Good Results: The following table shows results obtained for a CaO material using a Spectro Analytical Instruments Halogen (UV Plus) spectrometer. This test was done primarily to check delivery smoothness of the PowderPump™, taking five (5) replicate measurements over perhaps sixty (60) seconds. Better RSD's (relative standard deviations) were likely possible by integrating data collection over the entire sample delivery and normalizing the data to the mass of sample delivered. (Mass of sample may be determined by measuring mass of vial before and after sampling, giving the mass of sample delivered to the plasma.) Note Cl, Br,I in the results:

Spectro
CaO Li Be B Na Mg Si P S Cl K Ca Sc
Avg.(cps) 7228 346.3 1875 297081 2048100 2682800 1800 118384 254.6 259861 3074900 466.9
SD (n=5) 312.9 5.28 74.5 18506 182089 26511376.5 8844 8.62 108244643 14.61
%RSD 4.329 1.524 3.973 6.2298.891 9.882 4.251 7.471 3.384 4.1650.151 3.129














V Cr MnFe Co Ni Cu Zn As As_1Se Br
Avg.(cps) 790 33556 10836 530091329 1625 496538 18130 537 260.2344.6 10684
SD (n=5) 37.03 4095 778 233249.18 128.5 32868 1240 23.199.54 20.98 1288
%RSD 4.687 12.204 7.184 4.3993.7 7.906 6.619 6.839 4.319 3.6676.089 12.053


























Avg.(cps) MoAg Cd Sn Sb IBa W Tl Pb Pb_1 Bi
SD (n=5) 698 323.8 783 1391727 452.3 9010 1262 703 2824648 2026
%RSD 30.21 55.2 45.34 195.438.24 42.22 114576.9 35.85 391.367.9 118.9

4.327 17.042 5.788 14.0485.258 9.334 12.705 6.093 5.102 13.86 10.469 5.765